New 7310 Visible Spectrophotometer from Jenway

Select Country


To continue: Please select the country where you are delivering to.



7310 Visible Single Beam Spectrophotometer / 731001

• Icon driven software
• Results and method saving to USB memory stick
• Autologging capabilities
• 3 year warranty
• Extensive range of accessories available

The 7310 spectrophotometer uses icon driven software and has an improved navigation system for easy and intuitive usability. This spectrophotometer features measurement modes for absorbance, % transmittance, concentration, spectrum scanning, kinetics and quantitation.

The 7310 has an easy access USB port on the front of the unit which enables results to be stored directly to a USB memory stick for easy transfer of data. These models can also store up to 240 methods which can be copied from one unit to another using the USB capability.

The 7310 and 7315 models have enhanced security features enabling the settings and methods to be locked preventing unauthorised changes to the instrument settings and method parameters. The home page can also be customised and selected modes restricted, which is ideal for teaching laboratories.

The 73 series spectrophotometers are all covered by a three year warranty.

The 7310 visible, scanning spectrophotometer is fitted with a 10x10mm cuvette holder and is supplied with a pack of 100 disposable cuvettes, instruction manual, universal power supply, USB memory stick, PC software on CD ROM and interface cable.


 



Quantity Spin UpSpin Down

IN STOCK!
7310 Visible Single Beam Spectrophotometer


Full product specification.move

Full product specification.

 

Parameter

7310

Wavelength Range

320 to 1000nm

Light source

Tungsten halogen lamp

Resolution

1nm

Accuracy

± 2nm

Repeatibility

± 0.5nm

Spectral bandwidth

5nm

Photometric Transmittance

0 to 199.9%  

Photometric Absorbance

-0.300 to 2.500

Photometric Accuracy

± 1%T, ± 0.01Abs at 1.000 Absorbance

Photometric Resolution

0.1%T, 0.001A

Concentration Range

-300 to 9999

Concentration Resolution

selec table 1/ 0.1/ 0.01/ 0.001  

Concentation Calibration

Blank with a single standard or factor

Concentration Units

no units, %, ppm, EBC, SRM, mEq/l, mEq, M, mM, µM, nM, U, U/l, U/ml, g/l, mg/l, µg/l, ng/l, g/dl, mg/dl, µg/dl, mg/ml, µg/ml, ng/ml, µg/ul, ng/ul, mol/l, mmol/l

Concentration Factor

0.001 to 10000  

Concentration Standard

0.001 to 1000  

Quantitation Range

-300 to 9999

Quantitation Resolution

selectable 1/ 0.1/ 0.01/ 0.001  

Quantitation Calibration

Blank with up to 6 standards

Quantitation Curve fit algorithms

Quadratic, quadratic through zero, linear, linear through zero, interpolate  

Kinetics Measurement time

2 to 9999 seconds

Kinetics Calibration

Blank with a single standard or factor

Kinetics Display

Graphical and concentration  

Kinetics Analysis

Concentration, rate of change, initial and final absorbance/%T  

Kinetics Resolution

selectable 1/ 0.1/ 0.01/ 0.001  

Spectrum Range

320 to 1000nm  

Spectrum Scan interval

selectable 1, 2 or 5nm  

Spectrum Analysis

Absorbance or transmittance and peak and valleys  

GLP

Current time and date, user ID, instrument lock and method lock  

Method storage

240  

Results memory

Limited by attached mass storage device  

Removable media

USB (supplied)  

Outputs

USB, Analogue, RS232, Integral printer  

Size (w x d x h)

275 x 400 x 220mm  

Weight

6kg